Fluke 7526A: (1 yr) VER RS-232 /5520,8508+PRT,742A-1

Date: 2013-07-02 12:38:59

Revision: 3100

Automated Reference: Fluke 5520A

Automated Reference: Fluke 8508A

Manual Reference: Fluke 742A-1

Manual Reference: Fluke 8508A-PRT

Model(s) Covered:

Fluke 7526A

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1, based on using a 5520A characterized using a Fluke 8508A

(1 year) specifications.

THERMOCOUPLE SOURCE

- 75 mV Range, 10.000 mV; TUR 3.38

- 75 mV Range, 15.000 mV; TUR 3.45

- 75 mV Range, 35.000 mV; TUR 3.67

- 75 mV Range, 55.000 mV; TUR 3.84

- 75 mV Range, 75.000 mV; TUR 3.97

OHMS SOURCE

- 400 Ohm Range, 200 Ohms; TUR 3.06

- 400 Ohm Range, 300 Ohms; TUR 2.52

- 400 Ohm Range, 400 Ohms; TUR 2.14

OHMS MEASUREMENT

- 400 Ohm Range, 25 Ohms; TUR 2.73

- 400 Ohm Range, 75 Ohms; TUR 2.75

- 400 Ohm Range, 95 Ohms; TUR 2.83

- 400 Ohm Range, 100 Ohms; TUR 2.84

- 400 Ohm Range, 105 Ohms; TUR 2.26

- 400 Ohm Range, 150 Ohms; TUR 2.43

- 400 Ohm Range, 300 Ohms; TUR 2.74

- 400 Ohm Range, 400 Ohms; TUR 2.82

- 4000 Ohm Range, 1500 Ohms; TUR 2.62

- 4000 Ohm Range, 2000 Ohms; TUR 2.57

- 4000 Ohm Range, 3000 Ohms; TUR 2.74

- 4000 Ohm Range, 4000 Ohms; TUR 2.82

ISOLATED DC VOLTAGE

- 100 V Range , 75 Volts ; TUR 3.83

- 100 V Range , 100 Volts; TUR 3.59

Source:

Fluke 7526A Calibration Manual March 2013 *

* The tolerances used by this procedure were taken from the

published specifications table.

The average run time during testing was 37 minutes.