Fluke 753: (2 yr) VER RS-232 /5522,8508,LFCTR,1560+2565

Date: 2013-06-19 10:00:22

Revision: 2878

Automated Reference: Fluke 1560

Automated Reference: Fluke 5522A

Automated Reference: Fluke 8508A

Automated Reference: Low Frequency Counter

Manual Reference: Fluke 2565

Model(s) Covered:

753

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1:

DC CURRENT MEASURE using 5522A 1 year specifications

±30 mA; TUR 2.6786

±100 mA; TUR 3.7857

TC MEASUREMENT (Type K) using 5522A 1 year specifications

-180 °C; TUR 3.6364

0 °C; TUR 3.7500

1300 °C; TUR 3.0000

Source:

Fluke 753/754 Documenting Process Calibrator

Calibration Manual

November 2011

Compatibility:

MET/CAL 8.0.42 or later

Additional Equipment Required:

USB to RS-232 Plug-n-Play adapter, or

DB9 (f) to UUT LEMO pressure port cable assembly Fluke PN 949466

Dual Banana to Dual Banana cable (2 ea)

Single Banana to Single Banana cable (3 ea)

Type K Thermocouple Extension (1 ea)

Flexible Standard(s) Required:

See "Sub Fluke 75x Additional Required Equipment /5522,2565"

The average run time during testing was: 24 minutes (TAR mode).

The average run time during testing was: 1 hour, 10 minutes (MU mode).