Fluke 754: (2 yr) VER RS-232 /5522,8508,LFCTR,1560+2565, Revision: 1.4
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 754: (2 yr) VER RS-232 /5522,8508,LFCTR,1560+2565
Date: 2012-01-26 11:34:10
Revision: 1.4
Automated Reference: Fluke 1560
Automated Reference: Fluke 5522A
Automated Reference: Fluke 8508A
Automated Reference: Low Frequency Counter
Automated Reference: Low-Frequency Counter
Manual Reference: Fluke 2565
Model(s) Covered:
754
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1:
DC CURRENT MEASURE using 5522A 1 year specifications
±30 mA; TUR 2.6786
±100 mA; TUR 3.7857
TC MEASUREMENT (Type K) using 5522A 1 year specifications
-180 °C; TUR 3.0303
0 °C; TUR 2.5000
1300 °C; TUR 2.5000
TC SOURCE (Type K) using 2565 1 year specifications
-180 °C; TUR 3.9735
Source:
Fluke 753/754 Documenting Process Calibrator
Calibration Manual
November 2011
Compatibility:
MET/CAL 8.0.42 or later
Additional Equipment Required:
USB to RS-232 Plug-n-Play adapter, or
DB9 (f) to UUT LEMO pressure port cable assembly Fluke PN 949466
Rosemount 644H and programming cable
Dual Banana to Dual Banana cable (2 ea)
Single Banana to Single Banana cable (3 ea)
Type K Thermocouple Extension (1 ea)
Flexible Standard(s) Required:
See "Sub Fluke 75x Additional Required Equipment /5522,2565"
The average run time during testing was: 24 minutes (TAR mode).