Package: fluke_772_773_cal_9100_dmm.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 772: (1 year) CAL VER /9100,DMM, Revision: 1.1
Fluke 773: (1 year) CAL VER /9100,DMM, Revision: 1.1
Fluke 772: CAL ADJ /9100,DMM, Revision: 1.1
Fluke 773: CAL ADJ /9100,DMM, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 772: (1 year) CAL VER /9100,DMM
Fluke 773: (1 year) CAL VER /9100,DMM
Date: 2009-10-13 07:20:34
Revision: 1.1
Automated Reference: Datron 9100
Automated Reference: Digital Multimeter
Date: 2009-10-13 07:20:39
Revision: 1.1
Automated Reference: Datron 9100
Automated Reference: Digital Multimeter
Model(s) Covered:
772, 773
Source:
Fluke 772/773 Milliamp Process Clamp Meters Service Information.
January 2004
Source Author:
Fluke
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Cable, 14 gauge insulated copper
Flexible Standard(s) Required:
Digital Multimeter (DMM)
DC Voltage Range: 10 V
Accuracy: ±0.01% or better
DC Current Range: 24 mA
Accuracy: ±0.01% or better
Example: Fluke 8845A or equivalent
The average run time during testing was 6 minutes or less.
Model(s) Covered:
772, 773
Source:
Fluke 772/773 Milliamp Process Clamp Meters Service Information.
January 2004
Source Author:
Fluke
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Cable, 14 gauge insulated copper
Flexible Standard(s) Required:
Digital Multimeter (DMM)
DC Voltage Range: 10 V
Accuracy: ±0.01% or better
DC Current Range: 24 mA
Accuracy: ±0.01% or better
Example: Fluke 8845A or equivalent
The average run time during testing was 6 minutes or less.