Package: fluke_789_1yr_ver_9100_dmm.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 789: (1 year) CAL VER RS-232 /9100,DMM, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 789: (1 year) CAL VER RS-232 /9100,DMM
Date: 2010-11-24 07:03:43
Revision: 1.2
Automated Reference: Datron 9100
Automated Reference: Digital Multimeter
Model(s) Covered:
Fluke 789
Source:
Fluke 789 ProcessMeter Calibration Manual, Part No. 1988891,
September 2002, Manual Supplement 2, 5/04
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on the 5500A 1 year specifications.
CURRENT MEASUREMENT TEST (mA DC)
4.000 mA : TUR 2.7397
12.000 mA : TUR 3.1008
20.000 mA : TUR 3.2432
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
1 kOhm shunt, 1%, 2W, low TC preferable
IR Interface Adapter and IR Cable Adapter connector
The average run time during testing was 9 minutes.