Package: fluke_789_1yr_ver_9100_dmm.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 789: (1 year) CAL VER RS-232 /9100,DMM

Date: 2010-11-24 07:03:43

Revision: 1.2

Automated Reference: Datron 9100

Automated Reference: Digital Multimeter

Model(s) Covered:

Fluke 789

Source:

Fluke 789 ProcessMeter Calibration Manual, Part No. 1988891,

September 2002, Manual Supplement 2, 5/04

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on the 5500A 1 year specifications.

CURRENT MEASUREMENT TEST (mA DC)

4.000 mA : TUR 2.7397

12.000 mA : TUR 3.1008

20.000 mA : TUR 3.2432

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

1 kOhm shunt, 1%, 2W, low TC preferable

IR Interface Adapter and IR Cable Adapter connector

The average run time during testing was 9 minutes.