Package: fluke_8060a_1yr_ver_5500_5725.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 8060A: (1 year) CAL VER /5500,5725

Date: 2011-04-19 15:33:11

Revision: 1.7

Automated Reference: Fluke 5500A

Automated Reference: Fluke 5725A

Model(s) Covered:

8060A

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

0 Ohms; TUR 2.5000

100 Ohms; TUR 3.7500

40 MOhms; TUR 3.4063

Source:

Fluke 8060A instruction manual

P/N 632661, March 1982, Rev. 5, 1/93

Fluke 8060A PCB Replacement Kit

Installation Instruction Sheet

P/N 663564 March 2, 1999

Source Author:

Fluke

Compatibility:

MET/CAL or 5500/CAL 7.3 SP1 (7.3.36) or later

Additional Equipment Required:

None