Package: fluke_85_1yr_ver_5500.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 85: (1 year) CAL VER /5500

Date: 2011-11-11 13:39:50

Revision: 1.5

Automated Reference: Fluke 5500A

Model(s) Covered:

85

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

Conductance

10.00 nS; TUR 3.9365

DC Current

3500 mA DC; TUR 3.7037

AC Current

350 µA @ 60 Hz; TUR 3.5385

350 µA @ 1 kHz; TUR 3.5385

3500 µA @ 60 Hz; TUR 3.7398

3500 µA @ 1 kHz; TUR 3.7398

35 mA @ 60 Hz; TUR 3.7398

35 mA @ 1 kHz; TUR 3.7398

350 mA @ 60 Hz; TUR 3.5385

350 mA @ 1 kHz; TUR 3.5385

3500 mA @ 1 kHz; TUR 1.6974

10 A @ 1 kHz; TUR 2.2857

Capacitance

0.47 µF; TUR 3.5402

0.047 µF; TUR 3.5402

4.7 nF; TUR 2.2985

Source:

Fluke 83,84,87 Multimeters

Service Manual

PN 834168

April 1989

Rev 5, 3/96

Compatibility:

MET/CAL or 5500/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

None

The average run time during testing was 14 minutes (TAR mode).