Package: fluke_86x_rs232_5520_33250_8508.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 863: (1 year) CAL VER RS-232 /5520,33250

Date: 2007-04-16 15:29:43

Revision: 1.5

Automated Reference: Agilent 33250A

Automated Reference: Fluke 5520A

Model(s) Covered: 863

Source: Fluke 86X Performance Test Procedure, Preliminary,

February 1995

867B,863 Graphical Multimeters Service Manual

PN 689312 December 1997

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

AC Current

30 µA, 10 kHz; TUR 2.1053

2.9 A, 3 kHz; TUR 1.4710

1.0 A, 3 kHz; TUR 3.0000

2.9 A, 3 kHz; TUR 2.0645

TURs are for 5520A 90 day specifications.

Compatibility: MET/CAL or 5500/CAL 7.11b or later

Additional Equipment Required:

Fluke PM9080 Optically Isolated Interface Cable

Fluke Adapter Kit, Part No. 689486 or equivalent,

Includes DB9(f) to DB9(f) null modem adapter and

DB9(m) to DB9(m) gender changer

Cable, BNC(m)-(m)

BNC female to double stacking banana plug, Pomona 1269

The average run time during testing was 20 minutes.

Fluke 863: CAL ADJ RS-232 /5520

Fluke 865: CAL ADJ RS-232 /5520

Fluke 867: CAL ADJ RS-232 /5520

Fluke 867B: CAL ADJ RS-232 /5520

Date: 2007-04-16 15:30:07

Revision: 1.3

Automated Reference: Fluke 5520A

Model(s) Covered: 863, 865, 867, 867B

Source: Fluke 86X Calibration Procedure, Preliminary,

February 1995

867B,863 Graphical Multimeters Service Manual

PN 689312 December 1997

Compatibility: MET/CAL or 5500/CAL 7.11 or later

Additional Equipment Required:

Fluke PM9080 Optically Isolated Interface Cable

Power Adapter, BE860 (supplied with UUT)

Fluke Adapter Kit, Part No. 689486 or equivalent,

Includes DB9(f) to DB9(f) null modem adapter and

DB9(m) to DB9(m) gender changer

The average run time during testing was 23 minutes.

Fluke 865: (1 year) CAL VER RS-232 /5520,33250,8508

Fluke 867: (1 year) CAL VER RS-232 /5520,33250,8508

Fluke 867B: (1 year) CAL VER RS-232 /5520,33250,8508

Date: 2007-04-16 15:30:12

Revision: 1.6

Automated Reference: Agilent 33250A

Automated Reference: Fluke 5520A

Automated Reference: Fluke 8508A

Model(s) Covered: 865, 867, 867B

Source: Fluke 86X Performance Test Procedure, Preliminary,

February 1995

867B,863 Graphical Multimeters Service Manual

PN 689312 December 1997

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

AC Current

30 µA, 10 kHz; TUR 2.1053

2.9 A, 3 kHz; TUR 1.4710

1.0 A, 3 kHz; TUR 3.0000

2.9 A, 3 kHz; TUR 2.0645

TURs are for 5520A 90 day specifications.

Compatibility: MET/CAL 7.11b or later

Additional Equipment Required:

Fluke PM9080 Optically Isolated Interface Cable

Fluke Adapter Kit, Part No. 689486 or equivalent,

Includes DB9(f) to DB9(f) null modem adapter and

DB9(m) to DB9(m) gender changer

Cable, BNC(m)-(m)

BNC female to double stacking banana plug, Pomona 1269

The average run time during testing was 20 minutes.