Fluke 87V/AN: (1 year) CAL VER /9100, Revision: 2.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 87V/AN: (1 year) CAL VER /9100
Date: 2010-12-16 06:44:06
Revision: 2.1
Automated Reference: Datron 9100
Model(s) Covered:
87V/AN
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 9100 (1 year) specifications.
DC MILLIAMPERE TESTS (400 mA Range)
330 mA DC; TUR 2.8481
Source:
Fluke 87 V/AN Service Manual
PN 2153596, July 2004
Compatibility:
MET/CAL 7.20 SP3 (7.20Z) or later
Additional Equipment Required:
Probe, test, safety plug (normally supplied with UUT)
Fluke 80AK K-Type (thermocouple) DMM Adapter
Type K thermocouple extension cable with SMP (m)-(m) plugs
The average run time during testing was 16 minutes (keyboard entry).