Package: fluke_87_v_an_1yr_ver_9100.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Fluke 87V/AN: (1 year) CAL VER /9100

Date: 2010-12-16 06:44:06

Revision: 2.1

Automated Reference: Datron 9100

Model(s) Covered:

87V/AN

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 9100 (1 year) specifications.

DC MILLIAMPERE TESTS (400 mA Range)

330 mA DC; TUR 2.8481

Source:

Fluke 87 V/AN Service Manual

PN 2153596, July 2004

Compatibility:

MET/CAL 7.20 SP3 (7.20Z) or later

Additional Equipment Required:

Probe, test, safety plug (normally supplied with UUT)

Fluke 80AK K-Type (thermocouple) DMM Adapter

Type K thermocouple extension cable with SMP (m)-(m) plugs

The average run time during testing was 16 minutes (keyboard entry).