Fluke 87V EX: (1 year) CAL VER /9100

Date: 2010-12-16 06:44:13

Revision: 2.2

Automated Reference: Datron 9100

Model(s) Covered:

87V EX

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 9100 (1 year) specifications.

DC MILLIAMPERE TESTS (400 mA Range)

330 mA DC; TUR 2.8481

CAPACITANCE TESTS (10 nF Range)

4 nF; TUR 2.2222

Source:

Fluke 87V EX Calibration Manual

April 2006

Fluke 87V EX Calibration Manual Supplement

Issue Date: 12/10

Compatibility:

MET/CAL 7.20 SP3 (7.20Z) or later

Additional Equipment Required:

Probe, test, safety plug (normally supplied with UUT)

Fluke 80AK K-Type (thermocouple) DMM Adapter

Type K thermocouple extension cable with SMP (m)-(m) plugs

The average run time during testing was 16 minutes (keyboard entry).