Fluke 8808A: (1 year) CAL VER RS-232 /5520,8508

Date: 2010-10-14 07:07:07

Revision: 1.4

Automated Reference: Fluke 5520A

Automated Reference: Fluke 8508A

Manual Reference: Fluke 8508A

Model(s) Covered:

8808A

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1. In order to improve this, we use 8508A to characterize

the 5520A in advance for test points whose TUR which are lower than 3.5:1.

Here is list of TUR before (TUR1) and after characterization (TUR2).

DC CURRENT

- Range 200 µA: ±190 µA ; TUR1 1.38; TUR2 4.21

- Range 2000 µA: ±1900 µA ; TUR1 2.00; TUR2 8.98

- Range 200 mA: ±190 mA ; TUR1 3.40; TUR2 8.24

- Range 2 A: ±1.9 A ; TUR1 2.52; TUR2 5.48

- Range 10 A: ±10 A ; TUR1 3.82

AC CURRENT

- Range 200 mA: 190 mA, 2 kHz; TUR1 2.88; TUR2 8.77

- Range 2 A: 1.9 A, 2 kHz; TUR1 0.56; TUR2 3.69

Source:

8808A_Specs_F_Review.pdf

Testpoint_20071017.xls

Compatibility:

MET/CAL 7.20 SP1 or later

Additional Equipment Required:

4-Wire short, Fluke Part No. 2653346

The average run time during testing was 14 minutes in TAR mode.

The average run time during testing was 18 minutes in Meas. Unc. mode.