Fluke 902: (1 year) CAL VER /5520,5500A/COIL, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 902: (1 year) CAL VER /5520,5500A/COIL
Date: 2009-02-24 09:12:18
Revision: 1.1
Automated Reference: Fluke 5500A-COIL
Automated Reference: Fluke 5520A
Model(s) Covered:
902
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
CAPACITANCE:
90 µF, TUR 3.7624
1000 µF, TUR 3.8182
AC CURRENT:
0.4 A, 60 Hz TUR 2.4845
12 A, 60 Hz TUR 2.8808
12 A, 50 Hz TUR 2.8808
Source:
Fluke 902 HVAC CLAMP METER
Calibration Information, PN 2649733 July 2006
Source Author:
Fluke
Compatibility:
MET/CAL or 5500/CAL 7.20C or later
Additional Equipment Required:
TC Adapter Accessory
Type K thermocouple extension cable with SMP (m)-(m) plugs
(2) K-type mini-connectors
The average run time during testing was 10 minutes.