Package: fluke_902_1yr_ver_9100_opt200.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Fluke 902: (1 year) CAL VER /9100+Opt200, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Fluke 902: (1 year) CAL VER /9100+Opt200
Date: 2009-02-24 09:09:16
Revision: 1.1
Automated Reference: Datron 9100
Model(s) Covered:
902
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
CAPACITANCE:
90 µF, TUR 3.1148
1000 µF, TUR 3.1818
Source:
Fluke 902 HVAC CLAMP METER
Calibration Information, PN 2649733 July 2006
Source Author:
Fluke
Compatibility:
MET/CAL 7.20C or later
Additional Equipment Required:
TC Adapter Accessory
Type K thermocouple extension cable with SMP (m)-(m) plugs
(2) K-type mini-connectors
The average run time during testing was 10 minutes.