Package: general_resistance_rtd-100x_1yr_ver_8508.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

General Resistance RTD-100X: (1 year) CAL VER /8508

Date: 2011-09-26 10:38:43

Revision: 1.2

Automated Reference: Fluke 8508A

Models Covered:

RTD-100X

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 8508A 1 year specifications.

610.000 Ohms; TUR 3.9128

510.000 Ohms; TUR 3.7253

410.000 Ohms; TUR 3.4775

310.000 Ohms; TUR 3.1345

210.000 Ohms; TUR 2.6283

60.000 Ohms; TUR 3.8961

50.000 Ohms; TUR 3.7037

40.000 Ohms; TUR 3.4483

30.000 Ohms; TUR 3.0928

20.000 Ohms; TUR 2.5641

10.000 Ohms; TUR 3.7037

21.110 Ohms; TUR 2.6352

19.999 Ohms; TUR 2.5640

18.888 Ohms; TUR 3.9812

17.777 Ohms; TUR 3.9604

16.666 Ohms; TUR 3.9370

15.555 Ohms; TUR 3.9106

14.444 Ohms; TUR 3.8806

13.333 Ohms; TUR 3.8461

12.222 Ohms; TUR 3.8062

11.111 Ohms; TUR 3.7594

Source:

T.O.33K8-4-1136-1

TECHNICL MANUAL

CALIBRATION PROCEDURE FOR

RESISTANCE TEMPERATURE DETECTOR SIMULATOR

RTD-100X

28 FEBRUARY 2002

Compatibility:

MET/CAL 7.20B or later

Additional Equipment Required:

Low thermal EMF spade lug cables (2 ea.)

Fluke 5440A-7003 or equivalent

The average run time during testing was 17 minutes.