Gossen-Metrawatt SECUTEST SII N+:(1 yr)VER/5320+VLC,DMM, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Gossen-Metrawatt SECUTEST SII N+:(1 yr)VER/5320+VLC,DMM
Date: 2011-09-06 07:48:58
Revision: 1.2
Automated Reference: Digital Multimeter
Automated Reference: Fluke 5320A
Model(s) Covered:
SECUTEST SII N+
Source:
GOSSEN-METRAWATT SECUTEST SII
Operating Instructions
3-349-223-03, 8/11.05
GOSSEN-METRAWATT SECUTEST SII
Factory Calibration Certificate
Compatibility:
MET/CAL 7.3 or later
Flexible Standard(s) Required:
Digital Multimeter (DMM):
4-Wire Resistance:
Accuracy: ±0.5%
AC Current:
Accuracy: ±0.5%
Example: Fluke 8845A or equivalent
Additional Equipment Required:
DMM
The average run time during testing was 25 minutes (MU mode).