HP 33120A: (1 yr) VER IEEE /8845,5790,PSA,LFCTR

Date: 2009-09-24 12:36:09

Revision: 1.1

Automated Reference: Agilent PSA

Automated Reference: Fluke 5790A

Automated Reference: Fluke 8845A

Automated Reference: LF Counter

Automated Reference: Low Frequency Counter

Model(s) Covered:

33120A

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

AC AMPLITUDE (using 8845A)

- 1.4 V, 1 kHz; TUR 3.6458

- 140 mV, 1 kHz; TUR 3.6458

- 18 mV, 1 kHz; TUR 3.5433

Source:

HP 33120A Function Generator / Arbitrary Waveform Generator Service

Guide, Part No. 33120-90017, EWdition 6, March 2002

Compatibility:

MET/CAL 7.20x or later

Additional Equipment Required:

Termination, feedthrough, BNC (f)-(m), 50 Ohms ± 0.1 Ohms, HP/Agilent

11048C or equivalent

Adapter, BNC (f) to double stacking banana plug, Pomona Model 1269

or equivalent

Adapter, Type N (m) to BNC (f), Pomona Model 3288 or equivalent

Attenuator, 20 dB, Type N (m)-(f), 50 Ohms, Agilent 8491A or

equivalent (optional distortion tests only)

The average run time during testing was 22 minutes. (in main only)