Package: hp_34401a_1yr_ver_rs232_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Agilent 34401A: (1 year) CAL VER RS-232 /5520

HP 34401A: (1 year) CAL VER RS-232 /5520

Date: 2012-03-06 16:37:36

Revision: 1.5

Automated Reference: Fluke 5520A

Model Covered:

34401A

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1 based on 5520A 90 day specifications.

DC Voltage

- 100 mV; TUR 3.4000

- ±10 V; TUR 3.3333

- 100 V; TUR 3.0909

- 1000 V; TUR 3.3333

AC Voltage

- 1 V, 20 Hz; TUR 3.0000

- 10 V, 10 Hz; TUR 2.8571

- 750 V, 1 kHz; TUR 3.4177

Frequency Response

- 1 V, 20 Hz; TUR 3.0000

Resistance

- 100 Ohms; TUR 3.8889

- 10 MOhms; TUR 3.5652

AC Current

- 1 A, 1 kHz; TUR 3.0435

- 2 A, 1 kHz; TUR 3.6923

Source:

HP 34401A Multimeter Service Guide

Part Number: 34401-90013, Edition 4, February 1996

Agilent 34401A 6 1/2 Digit Multimeter Service Guide

Part Number: 34401-90013, Seventh Edition, August 2007

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Additional Equipment Required:

Precision 4-wire short with standard ¾" banana spacing:

Keithley 8610 Low Thermal Shorting Plug or equivalent

50 Ohm BNC(m)-(m) cable, 1 set

BNC female to double stacking banana plug, Pomona 1269

The average run time during testing was:

22 minutes (Root Difference Square guardbanding method used (TAR))

38 minutes (Measurement Uncertainty guardbanding method used (MU))