Package: hp_34401a_cal_ieee_5700_5725_fgen.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Agilent 34401A: (1 year) CAL VER IEEE /5700,5725,FGEN, Revision: 1.3
HP 34401A: (1 year) CAL VER IEEE /5700,5725,FGEN, Revision: 1.3
Agilent 34401A: CAL ADJ IEEE /5700,5725, Revision: 1.9
HP 34401A: CAL ADJ IEEE /5700,5725, Revision: 1.9
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Agilent 34401A: (1 year) CAL VER IEEE /5700,5725,FGEN
HP 34401A: (1 year) CAL VER IEEE /5700,5725,FGEN
Date: 2011-04-28 07:25:28
Revision: 1.3
Automated Reference: Fluke 5700A
Automated Reference: Fluke 5725A
Automated Reference: Function Generator
Agilent 34401A: CAL ADJ IEEE /5700,5725
HP 34401A: CAL ADJ IEEE /5700,5725
Date: 2011-04-28 07:25:30
Revision: 1.9
Automated Reference: Fluke 5700A
Automated Reference: Fluke 5725A
Model Covered:
34401A
Source:
HP 34401A Multimeter Service Guide
Part Number: 34401-90013, Edition 4, February 1996
Agilent 34401A 6 1/2 Digit Multimeter Service Guide
Part Number: 34401-90013, Seventh Edition, August 2007
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Precision 4-wire short with standard ¾" banana spacing:
Keithley 8610 Low Thermal Shorting Plug or equivalent
The average run time during testing was 20 minutes.
Model Covered:
34401A
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1 based on 5700A & 5725A 90 day specifications.
AC Voltage
- 100 mV, 50 kHz; TUR 3.9535
- 10 V, 10 Hz; TUR 1.3846
AC Current
- 1 A, 1 kHz; TUR 2.0290
- 2 A, 1 kHz; TUR 3.5821
2-Wire Ohm
- 100 Ohm; TUR 3.7838
Source:
HP 34401A Multimeter Service Guide
Part Number: 34401-90013, Edition 4, February 1996
Agilent 34401A 6 1/2 Digit Multimeter Service Guide
Part Number: 34401-90013, Seventh Edition, August 2007
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Precision 4-wire short with standard ¾" banana spacing:
Keithley 8610 Low Thermal Shorting Plug or equivalent
50 Ohm feed through terminator, 1 set
(Only for generator having the fixed 50 Ohm impedance)
50 Ohm BNC(m)-(m) cable, 1 set
BNC female to double stacking banana plug, Pomona 1269
The average run time during testing was:
20 minutes (Root Difference Square guardbanding method used (TAR))
35 minutes (Measurement Uncertainty guardbanding method used (MU))