Package: hp_34970a_cal_ieee_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Agilent 34970A: (1 year) CAL VER IEEE /5520

HP 34970A: (1 year) CAL VER IEEE /5520

Date: 2009-07-28 11:23:18

Revision: 1.7

Automated Reference: Fluke 5520A

Models covered:

HP/Agilent 34970A w/34901A 20 Channel Multiplexer module

The following test points have a Test Uncertainty Ratio (TUR) of less than

4 to 1 based 5520A 90 day specifications.

DC VOLTS, 100 mV Range : ±100 mV : TUR 3.6000

DC VOLTS, 10 V Range : ±10 V : TUR 3.3333

DC VOLTS, 100 V Range : ±100 V : TUR 3.0909

OHMS 2 WIRE, 100 Ohm Range : 100 Ohms : TUR 3.8889

OHMS 2 WIRE, 10 MOhm Range : 10 MOhms : TUR 3.5652

OHMS 4 WIRE, 100 Ohm Range : 100 Ohms : TUR 3.8889

OHMS 4 WIRE, 10 MOhm Range : 10 MOhms : TUR 3.5652

AC VOLTS, 1 V Range : 10 V @ 20 Hz : TUR 3.3333

AC VOLTS, 10 V Range : 10 V @ 10 Hz : TUR 3.1746

AC CURRENT, 10 mA Range : 10 mA @ 1 kHz : TUR 2.5455

AC CURRENT, 1 A Range : 1 A @ 1 kHz : TUR 3.0435

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Required Equipment:

HP/Agilent 34901A 20 Channel Multiplexer module

Copper wire, insulated, ~15 cm long

Leads, Teflon®, twisted pair, shielded, 3 ea. sufficiently long to

connect UUT to source

The average run time during testing was:

15 minutes (Root Difference Square guardbanding method Mode (TAR))

28 minutes (Measurement Uncertainty guardbanding method Mode (MU))

Agilent 34970A: CAL ADJ IEEE /5520

HP 34970A: CAL ADJ IEEE /5520

Date: 2009-07-28 11:23:20

Revision: 1.6

Automated Reference: Fluke 5520A

Models covered:

HP/Agilent 34970A w/34901A 20 Channel Multiplexer module

Required Equipment:

HP/Agilent 34901A 20 Channel Multiplexer module

Copper wire, insulated, ~15 cm long

Leads, Teflon®, twisted pair, shielded, 3 ea. sufficiently long to

connect UUT to source

The average run time during testing was 15 minutes.