Package: hp_34970a_cal_ieee_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Agilent 34970A: (1 year) CAL VER IEEE /5520, Revision: 1.7
HP 34970A: (1 year) CAL VER IEEE /5520, Revision: 1.7
Agilent 34970A: CAL ADJ IEEE /5520, Revision: 1.6
HP 34970A: CAL ADJ IEEE /5520, Revision: 1.6
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Agilent 34970A: (1 year) CAL VER IEEE /5520
HP 34970A: (1 year) CAL VER IEEE /5520
Date: 2009-07-28 11:23:18
Revision: 1.7
Automated Reference: Fluke 5520A
Agilent 34970A: CAL ADJ IEEE /5520
Date: 2009-07-28 11:23:20
Revision: 1.6
Automated Reference: Fluke 5520A
Models covered:
HP/Agilent 34970A w/34901A 20 Channel Multiplexer module
Required Equipment:
HP/Agilent 34901A 20 Channel Multiplexer module
Copper wire, insulated, ~15 cm long
Leads, Teflon®, twisted pair, shielded, 3 ea. sufficiently long to
connect UUT to source
The average run time during testing was 15 minutes.
Models covered:
HP/Agilent 34970A w/34901A 20 Channel Multiplexer module
The following test points have a Test Uncertainty Ratio (TUR) of less than
4 to 1 based 5520A 90 day specifications.
DC VOLTS, 100 mV Range : ±100 mV : TUR 3.6000
DC VOLTS, 10 V Range : ±10 V : TUR 3.3333
DC VOLTS, 100 V Range : ±100 V : TUR 3.0909
OHMS 2 WIRE, 100 Ohm Range : 100 Ohms : TUR 3.8889
OHMS 2 WIRE, 10 MOhm Range : 10 MOhms : TUR 3.5652
OHMS 4 WIRE, 100 Ohm Range : 100 Ohms : TUR 3.8889
OHMS 4 WIRE, 10 MOhm Range : 10 MOhms : TUR 3.5652
AC VOLTS, 1 V Range : 10 V @ 20 Hz : TUR 3.3333
AC VOLTS, 10 V Range : 10 V @ 10 Hz : TUR 3.1746
AC CURRENT, 10 mA Range : 10 mA @ 1 kHz : TUR 2.5455
AC CURRENT, 1 A Range : 1 A @ 1 kHz : TUR 3.0435
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Required Equipment:
HP/Agilent 34901A 20 Channel Multiplexer module
Copper wire, insulated, ~15 cm long
Leads, Teflon®, twisted pair, shielded, 3 ea. sufficiently long to
connect UUT to source
The average run time during testing was:
15 minutes (Root Difference Square guardbanding method Mode (TAR))