HP 8560A: (1 yr) CAL VER /9640,4418,HFSG,LFCTR,SWPG

Date: 2007-10-01

Revision: 1.2

Automated Reference: Agilent E4418B

Automated Reference: Agilent E9304A

Automated Reference: Fluke 9640A

Automated Reference: Fluke 9640A-50

Automated Reference: HF Signal Generator

Automated Reference: High Frequency Sig Gen

Automated Reference: HP 8481A

Automated Reference: LF Counter

Automated Reference: Low Frequency Counter

Automated Reference: Sweep Generator

Manual Reference: HP 11667A

Model(s) Covered:

HP 8560A with options

Source(s):

T.O. 33K3-4-2910-1, Technical Manual, Calibration Procedure For

Spectrum Analyzer 8560A(), Change 1, 30 March 2006

HP 8560A Installation and Verification Guide, Part No. 08560-90066,

November 1991

Compatibility:

MET/CAL 7.20r or later

Flexible Standard(s) Required:

Synthesized Sweeper (SWPG):

Frequency Range: 10 MHz to 10 GHz, CW

Accuracy: Locked to 10 MHz External Reference (UUT)

Amplitude Range: -35 to +16 dBm

Accuracy: ±2 dB

Internal and External (power meter) Leveling capability

SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset

HF Signal Generator (HFSG):

Frequency Range: 9 MHz to 2.5 GHz

Accuracy: ±100 kHz; Lockable to 10 MHz External Reference (UUT)

Amplitude Range: -20 to +10 dBm

Accuracy: ±2 dB

Phase Noise:

< -92 dBc/Hz @ 10 kHz offset

<-106 dBc/Hz @ 30 kHz offset

<-116 dBc/Hz @ 100 kHz offset

LF Counter (LFCTR):

Frequency Range: 10 MHz

Accuracy: Based on external frequency standard

(External frequency reference must be configured)

Time Interval Range: 30 ms to 60 s

Accuracy: ±0.25% or better (internal frequency reference)

Frequency Standard:

Frequency Range: Suitable for use with configured LF counter

Accuracy: (standard) ±0.625 ppm or better

(Option 003) ±0.03 ppm or better

Additional Equipment Required:

Power Splitter, 1 kHz to 18 GHz, 6 dB insertion loss, output

tracking: <0.25 dB, equivalent output SWR: <1.22, HP/Agilent

11667A or equivalent

Attenuator, Fixed, 20 dB, Type "N", 50 Ohms, DC to 6.5 GHz,

HP/Agilent 8491A Option 020, 2 ea.

Attenuator, Fixed, 10 dB, Type "N", 50 Ohms, DC to 6.5 GHz,

HP/Agilent 8491A Option 010

Cable assy., 36", 3.5 mm (m)-(m), 50 Ohm, DC to 12.4 GHz, Maximum

SWR: <1.4, 2 ea., HP/Agilent 8120-4921 or equivalent, 2 ea.

Cable assy., 24", 3.5 mm (m)-(m), 50 Ohm, DC to 12.4 GHz, Maximum

SWR: <1.4, 2 ea., HP/Agilent 11500E or equivalent

Cable assy., 36", BNC(m)-(m), 50 Ohm, 5 ea.

Cable assy., 6", BNC (m)-(m), 50 Ohm, DC to 300 MHz, HP/Agilent

10502A or equivalent

Adapter, Type "N" (m) to (m), 50 Ohms, DC to 12.4 GHz,

Agilent 1250-1475 or equivalent

Adapter, Type "N" (m) to 3.5 (f), 50 Ohms, DC to 12.4 GHz,

Instrument Grade, Agilent 1250-1744 or equivalent

Adapter, Type "N" (f) to BNC (m), 50 Ohms, DC to 300 MHz, Precision,

Agilent 1250-1477 or equivalent

Adapter, BNC (m) to SMA (f), 50 Ohms, DC to 300 MHz, Agilent E9634A

or equivalent

Adapter, Type "N" (m) to BNC (f), 50 Ohms, Agilent 1250-0780 or

equivalent

Tee, BNC (f)-(m)-(f), 50 Ohms, Agilent 1250-0781 or equivalent

The average run time during testing was 1 hour, 46 minutes (TAR mode).