HP 8560A: (1 yr) CAL VER /9640,4418,HFSG,LFCTR,SWPG
Date: 2007-10-01
Revision: 1.2
Automated Reference: Agilent E4418B
Automated Reference: Agilent E9304A
Automated Reference: Fluke 9640A
Automated Reference: Fluke 9640A-50
Automated Reference: HF Signal Generator
Automated Reference: HP 8481A
Automated Reference: LF Counter
Automated Reference: Low-Frequency Counter
Automated Reference: Sweep Generator
Manual Reference: HP 11667A
Model(s) Covered:
HP 8560A with options
Source(s):
T.O. 33K3-4-2910-1, Technical Manual, Calibration Procedure For
Spectrum Analyzer 8560A(), Change 1, 30 March 2006
HP 8560A Installation and Verification Guide, Part No. 08560-90066,
November 1991
Compatibility:
MET/CAL 7.20r or later
Flexible Standard(s) Required:
Synthesized Sweeper (SWPG):
Frequency Range: 10 MHz to 10 GHz, CW
Accuracy: Locked to 10 MHz External Reference (UUT)
Amplitude Range: -35 to +16 dBm
Accuracy: ±2 dB
Internal and External (power meter) Leveling capability
SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset
HF Signal Generator (HFSG):
Frequency Range: 9 MHz to 2.5 GHz
Accuracy: ±100 kHz; Lockable to 10 MHz External Reference (UUT)
Amplitude Range: -20 to +10 dBm
Accuracy: ±2 dB
Phase Noise:
< -92 dBc/Hz @ 10 kHz offset
<-106 dBc/Hz @ 30 kHz offset
<-116 dBc/Hz @ 100 kHz offset
LF Counter (LFCTR):
Frequency Range: 10 MHz
Accuracy: Based on external frequency standard
(External frequency reference must be configured)
Time Interval Range: 30 ms to 60 s
Accuracy: ±0.25% or better (internal frequency reference)
Frequency Standard:
Frequency Range: Suitable for use with configured LF counter
Accuracy: (standard) ±0.625 ppm or better
(Option 003) ±0.03 ppm or better
Additional Equipment Required:
Power Splitter, 1 kHz to 18 GHz, 6 dB insertion loss, output
tracking: <0.25 dB, equivalent output SWR: <1.22, HP/Agilent
11667A or equivalent
Attenuator, Fixed, 20 dB, Type "N", 50 Ohms, DC to 6.5 GHz,
HP/Agilent 8491A Option 020, 2 ea.
Attenuator, Fixed, 10 dB, Type "N", 50 Ohms, DC to 6.5 GHz,
HP/Agilent 8491A Option 010
Cable assy., 36", 3.5 mm (m)-(m), 50 Ohm, DC to 12.4 GHz, Maximum
SWR: <1.4, 2 ea., HP/Agilent 8120-4921 or equivalent, 2 ea.
Cable assy., 24", 3.5 mm (m)-(m), 50 Ohm, DC to 12.4 GHz, Maximum
SWR: <1.4, 2 ea., HP/Agilent 11500E or equivalent
Cable assy., 36", BNC(m)-(m), 50 Ohm, 5 ea.
Cable assy., 6", BNC (m)-(m), 50 Ohm, DC to 300 MHz, HP/Agilent
10502A or equivalent
Adapter, Type "N" (m) to (m), 50 Ohms, DC to 12.4 GHz,
Agilent 1250-1475 or equivalent
Adapter, Type "N" (m) to 3.5 (f), 50 Ohms, DC to 12.4 GHz,
Instrument Grade, Agilent 1250-1744 or equivalent
Adapter, Type "N" (f) to BNC (m), 50 Ohms, DC to 300 MHz, Precision,
Agilent 1250-1477 or equivalent
Adapter, BNC (m) to SMA (f), 50 Ohms, DC to 300 MHz, Agilent E9634A
or equivalent
Adapter, Type "N" (m) to BNC (f), 50 Ohms, Agilent 1250-0780 or
equivalent
Tee, BNC (f)-(m)-(f), 50 Ohms, Agilent 1250-0781 or equivalent
The average run time during testing was 1 hour, 46 minutes (TAR mode).