Package: hp_8560e-ec_2yr_ver_9640-lpn_e4418_lfctr_lo_pgen_swpg_scope.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
AT 8560E: (2 yr) VER /9640-LPN,E4418,LFCTR,LO,SWPG, Revision: 2
AT 8560EC: (2 yr) VER /9640-LPN,E4418,LFCTR,LO,SWPG, Revision: 2
HP 8560E: (2 yr) VER /9640-LPN,E4418,LFCTR,LO,SWPG, Revision: 2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
AT 8560E: (2 yr) VER /9640-LPN,E4418,LFCTR,LO,SWPG
AT 8560EC: (2 yr) VER /9640-LPN,E4418,LFCTR,LO,SWPG
HP 8560E: (2 yr) VER /9640-LPN,E4418,LFCTR,LO,SWPG
Date: 2011-08-25 13:10:13
Revision: 2
Automated Reference: Agilent E4418B
Automated Reference: Agilent E9304A
Automated Reference: Fluke 9640A
Automated Reference: Fluke 9640A-50
Automated Reference: HP 8481A
Automated Reference: HP 8485A
Automated Reference: LF Counter
Automated Reference: Local Oscillator
Automated Reference: Low-Frequency Counter
Automated Reference: Pulse Generator
Automated Reference: Sweep Generator
Manual Reference: Agilent 11667A
Manual Reference: Agilent 87300B
Manual Reference: Bench Oscilloscope
Model(s) Covered:
HP/Agilent 8560E/EC with options
Source(s):
Calibration Guide, Agilent Technologies 8560 E-Series and EC-Series
Spectrum Analyzers, Volume I & II, Part No. 08560-90171, May 2010
T.O. 33K3-4-3032-1, Technical Manual, Calibration Procedure For
Spectrum Analyzer 8563E(), 30 January 2002, Change 7, 30 March 2007
Compatibility:
MET/CAL 7.3 SP1 (7.3.36) or later
Flexible Standard(s) Required:
Synthesized Sweep Generator (SWPG):
Frequency Range: 10 MHz to 9821.4 MHz, CW
Frequency Accuracy: Lockable to 10 MHz External Reference
Amplitude Range: -35 dBm to +16 dBm
Amplitude Accuracy: ±2 dB
Pulse Modulation: Externally modulated
Local Oscillator (LO):
Frequency Range: 2 GHz to 2.00005 GHz, CW
Frequency Accuracy: Lockable to 10 MHz External Reference
Amplitude Range: -35 dBm to +13 dBm
Amplitude Accuracy: ±2 dB
Pulse Generator (PGEN):
Frequency Range: 1 Hz to 144 kHz
Pulse Width: 200 ns to 500 ms
Amplitude Range: 0 V LOW to 5 V HIGH
Low Frequency Counter (LFCTR):
Channel A
Frequency Range: 10 MHz
Frequency Accuracy: ±0.030 ppm or better (standard)
±0.625 ppm or better (Option 103)
Channel C (UUT Option 007 only)
Frequency Range: 300 MHz
Frequency Accuracy: < 0.025%
Time Interval
Range: 30 ms to 100 s
Accuracy: < 0.25% or better
Frequency Reference:
Frequency Range: Suitable for use with configured LFCTR
Frequency Accuracy: ±0.030 ppm or better (standard UUT)
±0.625 ppm or better (Option 103)
Additional Equipment Required (or equivalent):
See "Sub HP 856x Add'l Equipment Minimum Use Requirements"
for specific equipment, adapters, and cables required.
The average run times during testing were:
TAR mode: 1 hour, 5 minutes
MU mode: 2 hours, 15 minutes