HP 8563A: (1 yr) VER /9640-LPN,E4418,LFCTR,LFSG,LO,SWPG

Date: 2011-08-19 07:45:29

Revision: 2

Automated Reference: Agilent E4418B

Automated Reference: Agilent E9304A

Automated Reference: Fluke 9640A

Automated Reference: Fluke 9640A-50

Automated Reference: HP 8481A

Automated Reference: HP 8485A

Automated Reference: LF Counter

Automated Reference: LF Signal Generator

Automated Reference: Local Oscillator

Automated Reference: Low-Frequency Counter

Automated Reference: Pulse Generator

Automated Reference: Sweep Generator

Manual Reference: Agilent 11667B

Model(s) Covered:

HP 8563A with options

Source(s):

HP 8563A Installation and Verification Guide, Part No. 08563-90051,

Third Edition, August 1992

T.O. 33K3-4-2827-1, Technical Manual, Calibration Procedure For

Spectrum Analyzer 8563A(), Change 8, 30 December 2006

Compatibility:

MET/CAL 7.3 SP1 (7.3.36) or later

Flexible Standard(s) Required:

Synthesized Sweep Generator (SWPG):

Frequency Range: 10 MHz to 26.5 GHz, CW

Frequency Accuracy: Lockable to 10 MHz External Reference

Amplitude Range: -35 dBm to +16 dBm

Amplitude Accuracy: ±2 dB

Pulse Modulation: Externally modulated

Low Frequency Signal Generator (LFSG):

Frequency Range: 9 MHz to 45 MHz, CW

Frequency Accuracy: Lockable to 10 MHz External Reference

Amplitude Range: -20 to +10 dBm

Amplitude Accuracy: N/A, externally leveled

Local Oscillator (LO):

Frequency Range: 2 GHz to 8.00005 GHz, CW

Frequency Accuracy: Lockable to 10 MHz External Reference

Amplitude Range: -35 dBm to +13 dBm

Amplitude Accuracy: ±2 dB

Pulse Generator (PGEN):

Frequency Range: 14.4 kHz to 144 kHz

Pulse Width: 200 ns

Amplitude Range: 0 V LOW to 5 V HIGH

Low Frequency Counter (LFCTR):

Frequency Range: 10 MHz

Frequency Accuracy: ±0.030 ppm or better (standard)

±0.625 ppm or better (Option 103)

Time Interval

Range: 30 ms to 60 s

Accuracy: < 0.25% or better

Frequency Reference:

Frequency Range: Suitable for use with configured LFCTR

Frequency Accuracy: ±0.030 ppm or better (standard)

±0.625 ppm or better (Option 103)

Additional Equipment Required (or equivalent):

See "Sub HP 856x Add'l Equipment Minimum Use Requirements"

for specific equipment, adapters, and cables required.

The average run times during testing were:

TAR mode: 2 hour, 7 minutes

MU mode: 3 hours, 45 minutes