AT 8563E: (2 yr) VER /9640-LPN,E4418,LFCTR,LO,SWPG

AT 8563EC: (2 yr) VER /9640-LPN,E4418,LFCTR,LO,SWPG

HP 8563E: (2 yr) VER /9640-LPN,E4418,LFCTR,LO,SWPG

Date: 2012-10-17 14:02:02

Revision: 2.5

Automated Reference: Agilent E4418B

Automated Reference: Agilent E4419B

Automated Reference: Fluke 9640A

Automated Reference: Fluke 9640A-50

Automated Reference: HP 8482A

Automated Reference: HP 8485A

Automated Reference: Local Oscillator

Automated Reference: Low Frequency Counter

Automated Reference: Pulse Generator

Automated Reference: Sweep Generator

Manual Reference: Agilent 11667B

Manual Reference: Bench Oscilloscope

Manual Reference: Directional Coupler

Model(s) Covered:

HP/Agilent 8563E/EC with options

Source(s):

Calibration Guide, Agilent Technologies 8560 E-Series and EC-Series

Spectrum Analyzers, Volume I & II, Part No. 08560-90171, May 2010

T.O. 33K3-4-3032-1, Technical Manual, Calibration Procedure For

Spectrum Analyzer 8563E(), 30 January 2002, Change 7, 30 March 2007

Compatibility:

MET/CAL 7.3 SP1 (7.3.36) or later

Flexible Standard(s) Required:

Synthesized Sweep Generator (SWPG):

Frequency Range: 10 MHz to 26.5 GHz, CW

Frequency Accuracy: Lockable to 10 MHz External Reference

Amplitude Range: -80 dBm to +16 dBm

Amplitude Accuracy: ±2 dB

Pulse Modulation: Externally modulated

Local Oscillator (LO):

Frequency Range: 2 GHz to 8.00005 GHz, CW

Frequency Accuracy: Lockable to 10 MHz External Reference

Amplitude Range: -35 dBm to +13 dBm

Amplitude Accuracy: ±2 dB

Pulse Generator (PGEN):

Frequency Range: 1 Hz to 144 kHz

Pulse Width: 200 ns to 500 ms

Amplitude Range: 0 V LOW to 5 V HIGH

Output Impedance: 50 Ohms

Low Frequency Counter (LFCTR):

Channel A

Frequency Range: 10 MHz

Frequency Accuracy: ±0.030 ppm or better (standard)

±0.625 ppm or better (Option 103)

Channel C (UUT Option 007 only)

Frequency Range: 300 MHz

Frequency Accuracy: < 0.025%

Time Interval

Range: 30 ms to 100 s

Accuracy: < 0.25% or better

Frequency Reference:

Frequency Range: Suitable for use with configured LFCTR

Frequency Accuracy: ±0.030 ppm or better (standard UUT)

±0.625 ppm or better (Option 103)

Additional Equipment Required (or equivalent):

See "Sub HP 856x Add'l Equipment Minimum Use Requirements"

for specific equipment, adapters, and cables required.

The average run times during testing were:

TAR mode: 1 hour, 42 minutes

MU mode: 3 hours, 30 minutes