HP 8568B: (6 mo) CAL VER /9640,4418,HFSG,LFCTR

Date: 2009-05-01 10:24:03

Revision: 1.1

Automated Reference: Agilent E4418B

Automated Reference: Fluke 9640A

Automated Reference: Fluke 9640A-50

Automated Reference: HF Signal Generator

Automated Reference: HP 8482A

Automated Reference: LF Counter

Automated Reference: Low Frequency Counter

Manual Reference: Agilent 11667A

Model(s) Covered:

8568B

Source(s):

Performance Tests and Adjustment Manual, HP 8568B Spectrum Analyzer,

HP Part No. 08568-90115, June 1993

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Flexible Standards Required:

HF Signal Generator (HFSG):

Frequency Range: 30 MHz

Accuracy: ±0.02% (Internal)

Amplitude Range: -35 to -10 dBm

Resolution: 0.1 dB

Phase Noise at 30 MHz:

<-118 dBc/Hz @ 10 kHz offset

<-124 dBc/Hz @ 1 MHz offset

Example: HP 8663A or equivalent

LF Counter (LFCTR):

Ch. A Frequency Range: 10 MHz

Accuracy: Based on external frequency standard

Time Interval Range: 20 ms to 50 s

Accuracy: ±2.5% or better (internal frequency reference)

Example: Fluke PM 6690/6xx or equivalent

Additional Equipment Required:

Frequency Standard, suitable for use with configured LF counter,

±0.025 ppm or better

Power Splitter, 1 - 1500 MHz, Output Tracking <0.2 dB; Agilent

11667A or equivalent

Termination, Type N (m), 50 Ohms, DC - 3 GHz; Agilent 909C or

equivalent

Filter, Low Pass, 300 MHz; Agilent 0955-0455 or equivalent

Adapter, Precision, Type N (f) to BNC (m), 50 Ohms; Agilent 1250-1477

or equivalent

Cable Assy., 36", 50 Ohms, BNC (m)-(m), 3 ea.; Pomona Model 2249-C-36

or equivalent

Cable Assy., 12", BNC (m)-(m), 50 Ohm, DC to 20 MHz; Pomona Model

2249-C-12 or equivalent

The average run times during testing were:

1 hr. 46 min. (TAR mode)