Package: hp_8590l_1yr_ver_8340_3335_8662_6680_e4418.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

HP 8590L: (1 yr) CAL VER /8340,3335,8662,6680,4418

Date: 2007-04-16 15:32:37

Revision: 1.1

Automated Reference: Agilent E4418B

Automated Reference: HP 3335A

Automated Reference: HP 8340

Automated Reference: HP 8482A

Automated Reference: HP 8662A

Automated Reference: Philips PM 6680

Manual Reference: HP 11667A

Model(s) Covered:

8590L

Source:

This procedure is based on tests performed in U.S.A.F. Technical

Manual, T.O. 33K3-4-3091-1, dated 30 December 2003

Compatibility:

MET/CAL 7.11 or later

Additional Equipment Required:

HP/Agilent E9304A Power Sensor

Power Splitter, 50 Ohms, 10 MHz to 18 GHz, Type "N", HP/Agilent

11667A or equivalent

Termination, BNC (m), 50 Ohms, DC-2 GHz, HP/Agilent 11593A or

equivalent

Cable, coaxial, 4", 50 Ohms, BNC (m)-(m)

Cable, coaxial, 24", 50 Ohms, BNC (m)-(m)

Cable, coaxial, 36", 50 Ohms, BNC (m)-(m), RG-223/U

Cable, coaxial, 24", 50 Ohms, Type "N", (m)-(m)

Adapter, APC 3.5 mm (f) to Type "N" (f)

Adapter, Type "N" (m) to BNC (m), 50 Ohm

Adapter, BNC (m) to Type "N" (f), 50 Ohm

Adapter, BNC (f) to Type "N" (m), 50 Ohm

The average run times during testing were:

22 minutes (TUR mode)

2 hours, 31 minutes (measurement uncertainty mode)