HP 8594E: (1 yr) LIMITED CAL VER IEEE /9640,HFCTR

Date: 2012-01-10 09:02:35

Revision: 1.7

Automated Reference: Fluke 9640A

Automated Reference: Fluke 9640A-50

Automated Reference: High Frequency Counter

Model(s) Covered:

HP/Agilent 8594E

Source(s):

Agilent 8590 Series Analyzers Calibration Guide, Part Number

08594-90106, April 2001.

Compatibility:

MET/CAL version 7.3 SP1 (7.3.36) or later

Flexible Standards Required:

High Frequency Counter (HFCTR):

Frequency Range: 300 MHz

Aging Rate: < 1.875 ppm/year (Standard Timebase)

< 0.033 ppm/year (Option 004)

Additional Equipment/Accessories Required (or equivalent):

Termination(s):

Type "N" (m), 50 Ohms, DC to 3 GHz, Pomona 4240-50

Cable(s):

8½", 50 Ohms, BNC (m)-(m), Agilent 8120-2682

6", 50 Ohms, BNC (m)-(m), RG-58/U

36", 50 Ohms, BNC (m)-(m), RG-58/U

Adapter(s):

Type "N" (m) to BNC (f), 50 Ohm, Pomona 3288

BNC tee (f)-(m)-(f), 50 Ohm, Pomona 3285

The average run times during testing were:

TAR mode: 50 minutes

MU mode: 1 hour, 37 minutes