HP 8594E (w/Opt): (1 yr) VER /9640,441x,ESA,HFCTR,SWPG, Revision: 1.6
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
HP 8594E (w/Opt): (1 yr) VER /9640,441x,ESA,HFCTR,SWPG
Date: 2012-01-10 09:11:30
Revision: 1.6
Automated Reference: Agilent E4418B
Automated Reference: Fluke 9640A
Automated Reference: Fluke 9640A-50
Automated Reference: High Frequency Counter
Automated Reference: HP 8485A
Automated Reference: Low Frequency Counter
Automated Reference: Low-Frequency Counter
Automated Reference: Sweep Generator
Manual Reference: Agilent 11667B
Manual Reference: Agilent 87300B
Manual Reference: Bench Function Generator
Manual Reference: Bench Oscilloscope
Model(s) Covered:
HP/Agilent 8594E (with Options)
Source(s):
Agilent 8590 Series Analyzers Calibration Guide, Part Number
08594-90106, April 2001.
Compatibility:
MET/CAL version 7.3 SP1 (7.3.36) or later
Flexible Standards Required:
Sweep Generator (SWPG):
Frequency Range: 2000 to 9842.8 MHz
Frequency Accuracy (CW): ± 0.02%
Level Range: -20 to +10 dBm
High Frequency Counter (HFCTR):
Frequency Range: 300 MHz
Aging Rate: < 1.875 ppm/year (Standard Timebase)
< 0.033 ppm/year (Option 004)
Low Frequency Counter (LFCTR):
Function: Time Interval A-B
Period Range: 1 µs to 65 ms
Accuracy: < 0.01% of reading + 0.2 µs
Digital Multimeter (DMM):
Frequency Range: 9 to 100 kHz
Input Resistance: >= 10 MOhm
Accuracy: ±10 mV (on 100 V Range)
Additional Equipment/Accessories Required (or equivalent):
Bench Function Generator (Fluke 291)
100 Hz, 5 Vpp, Squarewave, 50 Ohm output impedance
Bench Oscilloscope (Tektronix TDS 3054C)
Accuracy: ±2% (at 0.5 µs/div)
Cursor Delta Time: ±5%
Power Splitter:
50 kHz to 22 GHz, 6 dB (nominal) insertion loss, output tracking
error: < 0.25 dB, VSWR: < 1.22, Agilent 11667B
Directional Coupler:
1.7 GHz to 8 GHz, Coupling: 16 dB, Coupling Deviation: ±1 dB,
Flatness: < 0.75 dB, VSWR: < 1.45, insertion loss < 1.3 dB,
Agilent 87300B
Termination(s):
Type "N" (m), 50 Ohms, DC to 3 GHz, Pomona 4240-50
Cable(s):
36", 50 Ohms, 3.5 mm (m)-(m), 10 MHz to 26.5 GHz, Max SWR: 1.4,
Max insertion loss: 2 dB, Agilent 8120-4921
8½", 50 Ohms, BNC (m)-(m), Agilent 8120-2682
6", 50 Ohms, BNC (m)-(m), RG-58/U
36", 50 Ohms, BNC (m)-(m), RG-58/U
Adapter(s):
APC 3.5 (f)-(f), 50 Ohm, Agilent 5061-5311
APC 3.5 (m)-(m), 50 Ohm, Agilent 1250-1159
APC 3.5 (m) to Type "N" (m), 50 Ohm, Agilent 1250-1743
APC 3.5 (m) to Type "N" (f), 50 Ohm, Agilent 1250-1750
Type "N" (m) to BNC (f), 50 Ohm, Pomona 3288
BNC tee (f)-(m)-(f), 50 Ohm, Pomona 3285
The average run times during testing were:
TAR mode: 1 hour, 2 minutes
MU mode: 1 hour, 47 minutes