HP 8648A: (2 yr) VER IEEE /N5531,PSA,LFCTR,HFCTR

HP 8648B: (2 yr) VER IEEE /N5531,PSA,LFCTR,HFCTR

HP 8648C: (2 yr) VER IEEE /N5531,PSA,LFCTR,HFCTR

HP 8648D: (2 yr) VER IEEE /N5531,PSA,LFCTR,HFCTR

Date: 2009-09-16 11:58:46

Revision: 1.1

Automated Reference: Agilent EPM

Automated Reference: Agilent N5532A

Automated Reference: Agilent PSA

Automated Reference: High Freq Counter

Automated Reference: High Frequency Counter

Automated Reference: Low Freq Counter

Automated Reference: Low Frequency Counter

Model(s) Covered:

8648A/B/C/D

Source:

HP 8648A/B/C/D Operation and Service Guide, Part No. 08648-90048,

April 1999

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Flexible Standard(s) Required:

LF Counter (LFCTR):

Frequency Range: 500 kHz to 100 MHz

Accuracy: Based on external frequency standard

Example: Fluke PM 6690/6xx Ch A or equivalent

HF Counter (HFCTR):

Frequency Range: 500 MHz to 4 GHz

Accuracy: Based on external frequency standard

Example: Fluke PM 6690/6xx CH C or equivalent

Frequency Standard:

Frequency Range: Suitable for use with configured LF & HF counters

Accuracy: ±0.001 ppm or better

Additional Equipment Required:

Agilent N5532A-504 Sensor Module

Cable assy., 24", BNC(m)-(m), 50 Ohm

The average run times during testing were:

1 hour, 40 minutes (TUR mode)

? hours, ?? minutes (measurement uncertainty mode)