HP 8656A: (1 yr) CAL VER IEEE /8902,8903,8560, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
HP 8656A: (1 yr) CAL VER IEEE /8902,8903,8560
Date: 2008-06-02 10:59:12
Revision: 1.1
Automated Reference: HP 8560
Automated Reference: HP 8902A
Automated Reference: HP 8903B
Manual Reference: HP 11722A
Model(s) Covered:
HP 8656A
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
RESIDUAL FM
- 105 MHz Carrier; TUR 3.00
FM INDICATOR ACCURACY, Carrier Frequency 100 MHz
- 99 kHz FM Deviation; TUR 3.62
Source:
HP 8656A Operating and Service manual, Part No. 08656-90108,
May 1981, Section 4, Performance Tests
Additional test points taken from those specified in the HP 11801A
Signal Generator Calibration Software.
Compatibility:
MET/CAL 7.01 SP1 or later
Additional Equipment Required:
HP/Agilent 11722A Sensor Module
Cable assy., 12", BNC(m)-(m), 50 Ohm
Cable assy., 24", Type "N"(m)-(m), 50 Ohm
Cable assy., 36", BNC(m)-(m), 50 Ohm, 2 ea.
Adapter, Type "N"(f) to BNC(m)
Adapter, tee, coaxial, BNC(f)-(m)-(f)
The average run times during testing were:
51 minutes (TUR mode)
2 hours, 1 minute (measurement uncertainty mode)