HP 8656A: (1 yr) CAL VER IEEE /8902,8903,8560

Date: 2008-06-02 10:59:12

Revision: 1.1

Automated Reference: HP 8560

Automated Reference: HP 8902A

Automated Reference: HP 8903B

Manual Reference: HP 11722A

Model(s) Covered:

HP 8656A

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

RESIDUAL FM

- 105 MHz Carrier; TUR 3.00

FM INDICATOR ACCURACY, Carrier Frequency 100 MHz

- 99 kHz FM Deviation; TUR 3.62

Source:

HP 8656A Operating and Service manual, Part No. 08656-90108,

May 1981, Section 4, Performance Tests

Additional test points taken from those specified in the HP 11801A

Signal Generator Calibration Software.

Compatibility:

MET/CAL 7.01 SP1 or later

Additional Equipment Required:

HP/Agilent 11722A Sensor Module

Cable assy., 12", BNC(m)-(m), 50 Ohm

Cable assy., 24", Type "N"(m)-(m), 50 Ohm

Cable assy., 36", BNC(m)-(m), 50 Ohm, 2 ea.

Adapter, Type "N"(f) to BNC(m)

Adapter, tee, coaxial, BNC(f)-(m)-(f)

The average run times during testing were:

51 minutes (TUR mode)

2 hours, 1 minute (measurement uncertainty mode)