Package: ifr_2394_95_1yr_ver_9640_8560_437_lfctr_lo_swpg_ieee.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
IFR 2394: (1 yr) VER IEEE /9640,8560,437,LFCTR,LO,SWPG, Revision: 1.1
IFR 2395: (1 yr) VER IEEE /9640,8560,437,LFCTR,LO,SWPG, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
IFR 2394: (1 yr) VER IEEE /9640,8560,437,LFCTR,LO,SWPG
IFR 2395: (1 yr) VER IEEE /9640,8560,437,LFCTR,LO,SWPG
Date: 2007-11-05 09:33:24
Revision: 1.1
Automated Reference: Fluke 9640A
Automated Reference: Fluke 9640A-50
Automated Reference: HP 437B
Automated Reference: HP 8560
Automated Reference: LF Counter
Automated Reference: Local Oscillator
Automated Reference: Low-Frequency Counter
Automated Reference: Sweep Generator
Automated Reference: Weinschel 8310
Manual Reference: HP 8485A
Model(s) Covered:
2394, 2395
Source(s):
Operation Manual, IFR Spectrum Analyzer 2394/2395, Section 6,
Performance Tests
Compatibility:
MET/CAL 7.2 or later
Flexible Standard(s) Required:
Synthesized Sweeper (SWPG):
Frequency Range: 10 MHz to 26.5 GHz
Accuracy: ±0.25%
Amplitude Range: -20 dBm to +10 dB
Resolution: 0.02 dB
Accuracy : NA
Leveling Modes: Internal and External (power meter)
SSB Phase Noise: <-86 dBc/Hz @ 10 kHz offset
External Reference Output: 10 MHz
LF Counter (LFCTR):
Frequency Range: 10 MHz
Accuracy: Based on external frequency standard
(External frequency reference must be configured)
Local Oscillator (LO):
Frequency Range: 10 MHz to 26.5 GHz
Accuracy: ±0.25%
Amplitude Range: -30 dBm to 0 dB
Resolution: 0.02 dB
Accuracy : NA
SSB Phase Noise: <-86 dBc/Hz @ 10 kHz offset
External Reference Output: 10 MHz