Krohn-Hite 6610: (1 year) CAL VER IEEE /CH5500

Date: 2010-06-17 12:49:33

Revision: 1.1

Automated Reference: Clarke-Hess 5500

Model(s) Covered:

6610

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

AMPLITUDE/FREQUENCY RESPONSE TESTS

- 180°, Ref: 250 mV, Var: 250 mV, 10 kHz; TUR 3.3333

- 180°, Ref: 250 mV, Var: 250 mV, 40 kHz; TUR 3.3333

- 45°, Ref: 10 V, Var: 10 V, 10 kHz; TUR 3.3333

- 45°, Ref: 10 V, Var: 10 V, 40 kHz; TUR 3.3333

- 270°, Ref: 20 V, Var: 20 V, 10 kHz; TUR 3.3333

- 270°, Ref: 20 V, Var: 20 V, 40 kHz; TUR 3.3333

- 315°, Ref: 75 V, Var: 75 V, 10 kHz; TUR 3.3333

- 315°, Ref: 75 V, Var: 75 V, 40 kHz; TUR 3.3333

Source(s):

USAF T.O. 33K8-4-922-1, Technical Manual, Calibration Procedure

For Phasemeter, 6610, 6620, dated 30 December 2001

Krohn-Hite Model 6620A (6620) Precision Phasemeter, 10 Hz to 10 MHz,

Operating Manual

Compatibility:

MET/CAL ??? or later

Additional Equipment Required:

Cable, coaxial, BNC (m)-(m), 2 ea., matched length and impedance,

Tektronix 012-0482-00 or equivalent

The average run time during testing was 9 minutes.