Package: keithley_196_1yr_ver_ieee_5700_5725.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Keithley 196: (1 year) CAL VER IEEE /5700,5725

Date: 2010-06-17 14:12:46

Revision: 1.1

Automated Reference: Fluke 5700A

Automated Reference: Fluke 5725A

Model(s) Covered:

Keithley 196

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1. Guardbanding is used to satisfy points whose TUR is

lower than 3.5:1.

DC CURRENT

- Range ±300 mA: ±300 mA ; TUR 3.8095

- Range ±3 A: ±3.0 A ; TUR 1.7949

Compatibility:

MET/CAL 7.20A or later

Source(s):

Keithley Model 196 System DMM Instruction Manual, Fourth Printing

January 1992, Document Number: 196-901-01 Rev. D