Package: keithley_2000_cal_ieee_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Keithley 2000: (1 year) CAL VER IEEE /5520

Date: 2007-06-20 10:05:33

Revision: 1.28

Automated Reference: Fluke 5520A

Model(s) Covered:

2000

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

DC Voltage

+/-1 V; TUR 3.5455

+/-10 V; TUR 2.9167

+/-100 V; TUR 3.2121

+/-1000 V; TUR 3.6970

AC Voltage

750 V, 1 kHz; TUR 3.4865

DC Current

+/-2.9 A; TUR 3.9560

AC Current

1.0 A, 1 kHz; TUR 3.0435

2.9 A, 1 kHz; TUR 3.3424

Resistance

10 MOhm, 2-Wire; TUR 3.5652

100 MOhm, 2-Wire; TUR 3.5581

Source:

Keithley Model 2000 Multimeter Calibration Manual,

Document Number: 2000-905-01 Rev. D, fourth Printing December 2001

Compatibility:

MET/CAL 7.10 or later

Additional Equipment Required:

None

The average run time during testing was 8 minutes.

Keithley 2000: CAL ADJ IEEE /5520

Date: 2007-06-20 10:05:37

Revision: 1.16

Automated Reference: Fluke 5520A

Model(s) Covered:

2000

Source:

Keithley Model 2000 Multimeter Calibration Manual,

Document Number: 2000-905-01 Rev. D, fourth Printing December 2001

Compatibility:

MET/CAL 7.10 or later

Additional Equipment Required:

Precision 4-wire short with standard ¾" banana spacing:

Keithley 8610 Low Thermal Shorting Plug or equivalent

The average run time during testing was 15 minutes.