Keithley 2000: (1 year) CAL VER IEEE /5700,5725

Date: 2008-12-11 13:52:32

Revision: 1.20

Automated Reference: Fluke 5700A

Automated Reference: Fluke 5725A

Model(s) Covered:

2000

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1 based on the 5700A 90 day specifications.

AC Voltage

100 mV, 50 kHz; TUR 3.9535

AC Current

1 A @ 1 kHz; TUR 2.0290

2.19999 A @ 1 kHz; TUR 3.4694

Source:

Keithley Model 2000 Multimeter Calibration Manual,

Document Number: 2000-905-01 Rev. D, fourth Printing December 2001

Compatibility:

MET/CAL 7.10 or later

Additional Equipment Required:

None

The average run time during testing was 18 minutes.

Keithley 2000: CAL ADJ IEEE /5700

Date: 2008-12-11 13:52:35

Revision: 1.15

Automated Reference: Fluke 5700A

Model(s) Covered:

2000

Source:

Keithley Model 2000 Service Manual,

Document Number: 2000-902-01 Rev. C, Third Printing June 1999

Compatibility:

MET/CAL 7.10 or later

Additional Equipment Required:

Precision 4-wire short with standard ¾" banana spacing:

Keithley 8610 Low Thermal Shorting Plug or equivalent

The average run time during testing was 15 minutes.