Keithley 2001: (2 yr) VER IEEE/5720WB,25,FGEN,100X,GL1G

Date: 2011-05-25 11:43:50

Revision: 1.3

Automated Reference: Fluke 5720A

Automated Reference: Fluke 5725A

Automated Reference: Function Generator

Manual Reference: General Resistance RTD-100X

Manual Reference: Guildline 9334A-1G

Models covered:

2001

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5720A (1 year) specifications.

AC Voltage

190 mV @ 10 Hz; TUR 2.0000

190 mV @ 25 kHz; TUR 2.2523

190 mV @ 50 kHz; TUR 2.5946

1.9 V @ 10 Hz; TUR 2.3226

19 V @ 10 Hz; TUR 2.6290

750 V @ 25 kHz; TUR 3.1725

AC Current

190 µA @ 10 kHz; TUR 2.2172

Source(s):

Keithley Model 2001 Multimeter Specifications

Rev. H / February 2009

Keithley 2001 Multimeter

Calibration Manual

©1992, Keithley Instruments, Inc.

Document Number: 2001-905-01 Rev. G

Compatibility:

MET/CAL 7.3 or later

Flexible Standard(s) Required:

Function Generator: 1 Hz to 15 MHz, 5 Vpp into 50 Ohms, Sine CW

Frequency Accuracy: 5 ppm

Additional Equipment Required:

50 Ohm feedthrough (BNC)