Keithley 2002: (2 yr) VER IEEE/5720W,25,90,FG,742,1G,X

Date: 2011-08-19 11:19:07

Revision: 1.1

Automated Reference: Fluke 5720A

Automated Reference: Fluke 5725A

Automated Reference: Fluke 5790A

Automated Reference: Function Generator

Manual Reference: Fluke 742A-10k

Manual Reference: Fluke 742A-1k

Manual Reference: General Resistance RTD-100X

Manual Reference: Guildline 9334A-1G

Models covered:

2002

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5720A (1 year) specifications.

DC Voltage

±1.9 V; TUR 3.4048

AC Voltage

190 mV @ 10 Hz; TUR 3.4783

1.9 V @ 10 Hz; TUR 2.7895

1.9 V @ 25 kHz; TUR 3.8905

19 V @ 10 Hz; TUR 3.7895

190 V @ 10 Hz; TUR 3.2895

750 V @ 25 kHz; TUR 2.5356

Resistance

19 Ohms; TUR 2.2326

190 Ohms; TUR 2.9010

AC Current

190 µA @ 10 kHz; TUR 2.4257

Source(s):

Keithley 2002 Multimeter

Calibration Manual

©1992, Keithley Instruments, Inc.

Compatibility:

MET/CAL 7.3 or later

Flexible Standard(s) Required:

Function Generator: 1 Hz to 15 MHz, 5 Vpp into 50 Ohms, Sine CW

Frequency Accuracy: 5 ppm

Additional Equipment Required:

50 Ohm feedthrough (BNC)