Package: keithley_2010_1yr_ver_rs232_5700_5725_33250_5125.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Keithley 2010: (1 yr) VER RS-232 /5700,5725,33250,5125, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Keithley 2010: (1 yr) VER RS-232 /5700,5725,33250,5125
Date: 2010-08-17 14:02:49
Revision: 1.2
Automated Reference: Agilent 33250A
Automated Reference: Fluke 5700A
Automated Reference: Fluke 5725A
Manual Reference: Hart 5125
Model(s) Covered:
2010
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
DC Voltage
+/-100 mV; TUR 3.4483
+/-1 V; TUR 3.7500
AC Voltage
100 mV, 50 kHz; TUR 3.9535
AC Current
1.0 A, 1 kHz; TUR 2.0290
2.19999 A, 1 kHz; TUR 3.4694
Resistance
10 Ohm, 4-Wire; TUR 2.6071
100 Ohm, 4-Wire; TUR 3.8235
RTD PT385
600 °C; TUR 2.2318
Source:
Keithley Model 2010 Multimeter Service Manual,
Document Number: 2010-902-01 Rev. C, Third Printing September 1999
Compatibility:
MET/CAL 7.11b or later
Required Files:
5125.acc
Accuracy file for Hart 5125 RTD Simulator
Additional Equipment Required:
Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36
BNC female to double stacking banana plug, Pomona 1269
The average run time during testing was 18 minutes.