Keithley 2010: (1 year) CAL VER RS-232 /5700,5725,5520
Date: 2010-08-17 14:16:45
Revision: 1.1
Automated Reference: Fluke 5520A
Automated Reference: Fluke 5700A
Automated Reference: Fluke 5725A
Model(s) Covered:
2010
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
DC Voltage
+/-100 mV; TUR 3.4483
+/-1 V; TUR 3.7500
AC Voltage
100 mV, 50 kHz; TUR 3.9535
AC Current
1.0 A, 1 kHz; TUR 2.0290
2.19999 A, 1 kHz; TUR 3.4694
Resistance
10 Ohm, 4-Wire; TUR 2.6071
100 Ohm, 4-Wire; TUR 3.8235
RTD PT385
-190 °C; TUR 3.5000
0 °C; TUR 1.6000
600 °C; TUR 1.4000
Source:
Keithley Model 2010 Multimeter Service Manual,
Document Number: 2010-902-01 Rev. C, Third Printing September 1999
Compatibility:
MET/CAL 7.11b or later
Additional Equipment Required:
None
The average run time during testing was 18 minutes.