Keithley 2010: CAL ADJ RS-232 /5700, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Keithley 2010: CAL ADJ RS-232 /5700
Date: 2010-08-17 13:30:50
Revision: 1.1
Automated Reference: Fluke 5700A
Model(s) Covered:
2010
Source:
Keithley Model 2010 Multimeter Service Manual,
Document Number: 2010-902-01 Rev. C, Third Printing September 1999
Compatibility:
MET/CAL 7.10 or later
Additional Equipment Required:
Precision 4-wire short with standard ¾" banana spacing:
Keithley 8610 Low Thermal Shorting Plug or equivalent
The average run time during testing was 24 minutes.