Package: keithley_2015_1yr_ver_rs232_5520_ds360_8508.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Keithley 2015: (1 year) VER RS-232 /5520,DS360,8508

Date: 2011-08-19 16:40:54

Revision: 2

Automated Reference: Fluke 5520A

Automated Reference: Fluke 8508A

Automated Reference: SRS DS360

Model(s) Covered:

2015

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

DC Voltage

+/-1 V; TUR 3.5455

+/-10 V; TUR 2.9167

+/-100 V; TUR 3.2121

+/-1000 V; TUR 3.6970

AC Voltage

750 V, 1 kHz; TUR 3.4865

DC Current

+/-2.9 A; TUR 3.9560

AC Current

1.0 A, 1 kHz; TUR 3.0435

2.9 A, 1 kHz; TUR 3.3424

Resistance

10 MOhm, 2-Wire; TUR 3.5652

100 MOhm, 2-Wire; TUR 3.5581

Sine Wave Output

4 V, 1 kHz; TUR 1.4286

Source:

Keithley Model 2015 Service Manual,

Document Number: 2015-902-01 Rev. C, Third Printing June 1999

Model 2015/2015P THD Multimeter User¡¯s Manual,

Document Number: 2015-900-01 Rev. F, Sixth Printing, August 2003

Compatibility:

MET/CAL 7.11 SP1 (7.11j) or later

Required Files:

DS360.acc

Accuracy file for the SRS DS360

Additional Equipment Required:

Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36

BNC female to double stacking banana plug, Pomona 1269

The average run time during testing was 15 minutes.