Package: keithley_2015_2016_cal_ieee_5700_5725_33250_ds360_8508.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Keithley 2015-P:(1yr)VER IEEE/5700/25,33250,DS360,8508

Keithley 2015:(1yr)VER IEEE/5700/25,33250,DS360,8508

Date: 2008-12-12 10:05:07

Revision: 1.5

Automated Reference: Agilent 33250A

Automated Reference: Fluke 5700A

Automated Reference: Fluke 5725A

Automated Reference: Fluke 8508A

Automated Reference: SRS DS36

Model(s) Covered:

2015, 2015-P

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

AC Voltage

100 mV, 50 kHz; TUR 3.9535

AC Current

1 A @ 1 kHz; TUR 2.0290

2.19999 A @ 1kHz; TUR 3.4694

Source:

Keithley Model 2015 Service Manual,

Document Number: 2015-902-01 Rev. C, Third Printing June 1999

Model 2015/2015P THD Multimeter User¡¯s Manual,

Document Number: 2015-900-01 Rev. F, Sixth Printing, August 2003

Compatibility:

MET/CAL 7.11 SP1 (7.11j) or later

Required Files:

DS360.acc

SRS DS360 Accuracy File

Additional Equipment Required:

Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36

BNC female to double stacking banana plug, Pomona 1269

The average run time during testing was 18 minutes.

Keithley 2016-P:(1yr)VER IEEE/5700/25,33250,DS360,8508

Keithley 2016:(1yr)VER IEEE/5700/25,33250,DS360,8508

Date: 2008-12-12 10:05:09

Revision: 1.5

Automated Reference: Agilent 33250A

Automated Reference: Fluke 5700A

Automated Reference: Fluke 5725A

Automated Reference: Fluke 8508A

Automated Reference: SRS DS36

Model(s) Covered:

2016, 2016-P

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

AC Voltage

100 mV, 50 kHz; TUR 3.9535

AC Current

1 A @ 1 kHz; TUR 2.0290

2.19999 A @ 1kHz; TUR 3.4694

Source:

Keithley Model 2016 Service Manual,

Document Number: 2016-902-01 Rev. A, First Printing May 1999

Model 2016/2016P THD Multimeter User¡¯s Manual,

Document Number: 2016-900-01 Rev. C, Third Printing, August 2003

Compatibility:

MET/CAL 7.11 SP1 (7.11j) or later

Required Files:

DS360.acc

SRS DS360 Accuracy File

Additional Equipment Required:

Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36

BNC female to double stacking banana plug, Pomona 1269

The average run time during testing was 25 minutes.

Keithley 2015-P: CAL ADJ IEEE /5700,DS360

Keithley 2015: CAL ADJ IEEE /5700,DS360

Keithley 2016-P: CAL ADJ IEEE /5700,DS360

Keithley 2016: CAL ADJ IEEE /5700,DS360

Date: 2008-12-12 10:05:12

Revision: 1.3

Automated Reference: Fluke 5700A

Automated Reference: SRS DS36

Automated Reference: SRS DS360

Model(s) Covered:

2015, 2015-P, 2016, 2016-P

Source:

Keithley Model 2015 Service Manual,

Document Number: 2015-902-01 Rev. C, Third Printing June 1999

Keithley Model 2016 Service Manual,

Document Number: 2016-902-01 Rev. A, First Printing May 1999

Compatibility:

MET/CAL 7.10 or later

Additional Equipment Required:

Precision 4-wire short with standard ¾" banana spacing:

Keithley 8610 Low Thermal Shorting Plug or equivalent

Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36

BNC female to double stacking banana plug, Pomona 1269

The average run time during testing was 15 minutes.