Keithley 2015-P:(1yr)VER RS232/5700/25,33250,DS360,8508, Revision: 1.6
Keithley 2015:(1yr)VER RS232/5700/25,33250,DS360,8508, Revision: 1.6
Keithley 2016-P:(1yr)VER RS232/5700/25,33250,DS360,8508, Revision: 1.6
Keithley 2016:(1yr)VER RS232/5700/25,33250,DS360,8508, Revision: 1.6
Keithley 2015-P: CAL ADJ RS-232 /5700,DS360, Revision: 1.3
Keithley 2015: CAL ADJ RS-232 /5700,DS360, Revision: 1.3
Keithley 2016-P: CAL ADJ RS-232 /5700,DS360, Revision: 1.3
Keithley 2016: CAL ADJ RS-232 /5700,DS360, Revision: 1.3
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Keithley 2015-P:(1yr)VER RS232/5700/25,33250,DS360,8508
Keithley 2015:(1yr)VER RS232/5700/25,33250,DS360,8508
Date: 2008-12-12 10:07:54
Revision: 1.6
Automated Reference: Agilent 33250A
Automated Reference: Fluke 5700A
Automated Reference: Fluke 5725A
Automated Reference: Fluke 8508A
Automated Reference: SRS DS36
Automated Reference: SRS DS360
Keithley 2016-P:(1yr)VER RS232/5700/25,33250,DS360,8508
Keithley 2016:(1yr)VER RS232/5700/25,33250,DS360,8508
Date: 2008-12-12 10:07:57
Revision: 1.6
Automated Reference: Agilent 33250A
Automated Reference: Fluke 5700A
Automated Reference: Fluke 5725A
Automated Reference: Fluke 8508A
Automated Reference: SRS DS36
Automated Reference: SRS DS360
Model(s) Covered:
2016, 2016-P
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
AC Voltage
100 mV, 50 kHz; TUR 3.9535
AC Current
1 A @ 1 kHz; TUR 2.0290
2.19999 A @ 1kHz; TUR 3.4694
Source:
Keithley Model 2016 Service Manual,
Document Number: 2016-902-01 Rev. A, First Printing May 1999
Model 2016/2016P THD Multimeter User¡¯s Manual,
Document Number: 2016-900-01 Rev. C, Third Printing, August 2003
Compatibility:
MET/CAL 7.11 SP1 (7.11j) or later
Required Files:
DS360.acc
SRS DS360 Accuracy File
Additional Equipment Required:
Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36
BNC female to double stacking banana plug, Pomona 1269
The average run time during testing was 15 minutes.
Keithley 2015-P: CAL ADJ RS-232 /5700,DS360
Keithley 2015: CAL ADJ RS-232 /5700,DS360
Keithley 2016-P: CAL ADJ RS-232 /5700,DS360
Keithley 2016: CAL ADJ RS-232 /5700,DS360
Date: 2008-12-12 10:08:00
Revision: 1.3
Automated Reference: Fluke 5700A
Automated Reference: SRS DS360
Model(s) Covered:
2015, 2015-P, 2016, 2016-P
Source:
Keithley Model 2015 Service Manual,
Document Number: 2015-902-01 Rev. C, Third Printing June 1999
Keithley Model 2016 Service Manual,
Document Number: 2016-902-01 Rev. A, First Printing May 1999
Compatibility:
MET/CAL 7.10 or later
Additional Equipment Required:
Precision 4-wire short with standard ¾" banana spacing:
Keithley 8610 Low Thermal Shorting Plug or equivalent
Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36
BNC female to double stacking banana plug, Pomona 1269
The average run time during testing was 15 minutes.
Model(s) Covered:
2015, 2015-P
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
AC Voltage
100 mV, 50 kHz; TUR 3.9535
AC Current
1 A @ 1 kHz; TUR 2.0290
2.19999 A @ 1kHz; TUR 3.4694
Sine Wave Output
9V, 1kHz; TUR 1.4286
Source:
Keithley Model 2015 Service Manual,
Document Number: 2015-902-01 Rev. C, Third Printing June 1999
Model 2015/2015P THD Multimeter User¡¯s Manual,
Document Number: 2015-900-01 Rev. F, Sixth Printing, August 2003
Compatibility:
MET/CAL 7.11 SP1 (7.11j) or later
Required Files:
DS360.acc
SRS DS360 Accuracy File
Additional Equipment Required:
Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36
BNC female to double stacking banana plug, Pomona 1269
The average run time during testing was 15 minutes.