Keithley 2016: (1 year) VER RS-232 /5520,DS360,8508

Date: 2011-08-19 16:52:45

Revision: 2

Automated Reference: Fluke 5520A

Automated Reference: Fluke 8508A

Automated Reference: SRS DS360

Model(s) Covered:

2016

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

DC Voltage

+/-1 V; TUR 3.5455

+/-10 V; TUR 2.9167

+/-100 V; TUR 3.2121

+/-1000 V; TUR 3.6970

AC Voltage

750 V, 1 kHz; TUR 3.4865

DC Current

+/-2.9 A; TUR 3.9560

AC Current

1.0 A, 1 kHz; TUR 3.0435

2.9 A, 1 kHz; TUR 3.3424

Resistance

10 MOhm, 2-Wire; TUR 3.5652

100 MOhm, 2-Wire; TUR 3.5581

Source:

Keithley Model 2016 Service Manual,

Document Number: 2016-902-01 Rev. A, First Printing May 1999

Model 2016/2016P THD Multimeter User¡¯s Manual,

Document Number: 2016-900-01 Rev. C, Third Printing, August 2003

Compatibility:

MET/CAL 7.11 SP1 (7.11j) or later

Required Files:

DS360.acc

Accuracy file for the SRS DS360

Additional Equipment Required:

Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36

BNC female to double stacking banana plug, Pomona 1269

The average run time during testing was 15 minutes.