Package: keithley_2016_1yr_ver_rs232_5520_ds360_8508.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Keithley 2016: (1 year) VER RS-232 /5520,DS360,8508, Revision: 2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Keithley 2016: (1 year) VER RS-232 /5520,DS360,8508
Date: 2011-08-19 16:52:45
Revision: 2
Automated Reference: Fluke 5520A
Automated Reference: Fluke 8508A
Automated Reference: SRS DS360
Model(s) Covered:
2016
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
DC Voltage
+/-1 V; TUR 3.5455
+/-10 V; TUR 2.9167
+/-100 V; TUR 3.2121
+/-1000 V; TUR 3.6970
AC Voltage
750 V, 1 kHz; TUR 3.4865
DC Current
+/-2.9 A; TUR 3.9560
AC Current
1.0 A, 1 kHz; TUR 3.0435
2.9 A, 1 kHz; TUR 3.3424
Resistance
10 MOhm, 2-Wire; TUR 3.5652
100 MOhm, 2-Wire; TUR 3.5581
Source:
Keithley Model 2016 Service Manual,
Document Number: 2016-902-01 Rev. A, First Printing May 1999
Model 2016/2016P THD Multimeter User¡¯s Manual,
Document Number: 2016-900-01 Rev. C, Third Printing, August 2003
Compatibility:
MET/CAL 7.11 SP1 (7.11j) or later
Required Files:
DS360.acc
Accuracy file for the SRS DS360
Additional Equipment Required:
Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36
BNC female to double stacking banana plug, Pomona 1269
The average run time during testing was 15 minutes.