Keithley 2110: (1 year) CAL VER IEEE /5700,5725,5502

Date: 2013-07-11 10:07:15

Revision: 3548

Automated Reference: Fluke 5502E

Automated Reference: Fluke 5700A

Automated Reference: Fluke 5725A

Model Covered:

2110

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1 using 5700A 1 year specifications.

AC Voltage

100 mV, 10 Hz; TUR 2.3944

1 V, 10 Hz; TUR 2.4286

10 V, 10 Hz; TUR 2.4286

100 V, 10 Hz; TUR 2.4286

DC Current

±1.5 A; TUR 3.8235

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1 using 5700A+5725A 1 year specifications.

AC Voltage

700 V, 75 kHz; TUR 3.1118

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1 using 5502E 1 year specifications.

Capacitance

1 nF; TUR 1.8667

100 µF; TUR 2.7273

Source:

Keithley Model 2110 5½ Digit Multimeter Calibration Manual,

Document Number: 2110-905-01 Rev. A / July 2012

Compatibility:

MET/CAL 8.1.4

The average run time during testing was 13 minutes.