Package: kyoritsu_20x7_1yr_ver_5520_55coil.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Kyoritsu KEW SNAP 2017: (1 year) VER /5520,5500A/COIL

Kyoritsu KEW SNAP 2027: (1 year) VER /5520,5500A/COIL

Date: 2007-03-23 14:06:41

Revision: 1.1

Automated Reference: Fluke 5500A-COIL

Automated Reference: Fluke 5520A

Model(s) Covered:

Kyoritsu KEW SNAP 2017, 2027 (True RMS)

This procedure contains the following test(s) whose Test Uncertainty Ratio

are less than 4:1.

AC CURRENT

- 20 A, 60 Hz; TUR 1.9324

- 190 A, 60 Hz; TUR 1.6575

- 400 A, 60 Hz; TUR 2.2192 (2017)

- 550 A, 60 Hz; TUR 2.2514 (2017)

- 400 A, 60 Hz; TUR 3.1313 (2027)

- 550 A, 60 Hz; TUR 2.9610 (2027)

Source:

Kyoritsu KEW SNAP 2027 Instruction Manual, Part No. 92-1281A,

Specification

Fluke Model 5500A/COIL Instruction Sheet, Part No. 600616, Aug. 1995,

Rev. 6, 12/02

Compatibility:

5500/CAL or MET/CAL 7.01 or later

Subprocedures:

None

Required Files:

None

Additional Equipment Required:

None

The average run time during testing was 8 minutes.

This procedure is intended for use with MET/CAL® calibration software;

the terms and conditions set forth in your MET/CAL license apply to this

procedure.

Due to Fluke's policy of continuously updating our products, this procedure

may contain minor differences in methods used and/or specifications to

those found in the manual or other documentation. While every effort has

been made to ensure that this procedure is accurate, Fluke cannot be held

responsible for the consequences of error or omissions found within this

procedure.