LeCroy WP7000: (1 yr) CAL VER IEEE /5520+G1,DMM

LeCroy WP7100: (1 yr) CAL VER IEEE /5520+G1,DMM

LeCroy WP7100A: (1 yr) CAL VER IEEE /5520+G1,DMM

Date: 2008-06-12 12:46:50

Revision: 1.1

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5520A

Models covered:

WavePro 7000, WavePro 7100(A)

Source(s):

Chapter 5 - WP7k Performance Verification revD.pdf

Chapter 5 - WP7k Performance Verification Test Record revD.pdf

LeCroy X-Stream Oscilloscope Operator's Manual, February 2007

- WM-OM-E_Rev_K.pdf

Compatibility:

MET/CAL 7.20 SP1 or later

Flexible Standard(s) Required:

Digital Multimeter (DMM)

Voltage Range: 100 mV

Accuracy: ±0.1%

Example: Fluke 8845A or equivalent

Additional Equipment Required:

Cable, 50 Ohm, BNC(m)-(m)

Adapter, BNC female to double stacking banana plug, Pomona 1269

or equivalent