LeCroy WR6030: (1 yr) CAL VER IEEE /5820

LeCroy WR6050: (1 yr) CAL VER IEEE /5820

LeCroy WR6051: (1 yr) CAL VER IEEE /5820

Date: 2008-10-07 15:13:45

Revision: 1.1

Automated Reference: Fluke 5800A

Model(s) Covered:

WR6030, 6050, and 6051

Source:

LeCroy WaveRunner® 6000 Series Service Manual, August 2004

Compatibility:

MET/CAL 7.20 or later

The average run time during testing was 16 minutes (2 channel),

31 minutes (4 channel).

LeCroy WR6030: (1 yr) CAL VER IEEE /5820+5ch

LeCroy WR6050: (1 yr) CAL VER IEEE /5820+5ch

LeCroy WR6051: (1 yr) CAL VER IEEE /5820+5ch

Date: 2008-10-07 15:13:50

Revision: 1.3

Automated Reference: Fluke 5800A

Automated Reference: Fluke 5820A

Model(s) Covered:

WaveRunner 6030, 6050, and 6051

Source:

LeCroy WaveRunner® 6000 Series Service Manual, August 2004

The following tests have a Test Accuracy Ratio (TAR) of less than 4:1.

LEAKAGE VOLTAGE TEST (all channels)

DC, 50 Ohms : 20 mV/div, TAR 2:1

DC, 50 Ohms : 200 mV/div, TAR 2:1

DC, 1 MOhm : 20 mV/div, TAR 2:1

DC, 1 MOhm : 200 mV/div, TAR 2:1

DC, 1 MOhm : 2 V/div, TAR 2:1

DC, 50 Ohms : External, TAR 2:1

DC, 1 MOhm : External, TAR 2:1

Compatibility:

MET/CAL 7.20 or later

The average run time during testing was 16 minutes (2 channel),

31 minutes (4 channel).