LeCroy WR6100: (1 yr) CAL VER IEEE /5800+GHz,DMM, Revision: 1.1
LeCroy WR6200: (1 yr) CAL VER IEEE /5800+GHz,DMM, Revision: 1.1
LeCroy WR6100: (1 yr) CAL VER IEEE /5800+GHz+5ch,DMM, Revision: 1.1
LeCroy WR6200: (1 yr) CAL VER IEEE /5800+GHz+5ch,DMM, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
LeCroy WR6100: (1 yr) CAL VER IEEE /5800+GHz,DMM
LeCroy WR6200: (1 yr) CAL VER IEEE /5800+GHz,DMM
Date: 2008-10-07 15:09:59
Revision: 1.1
Automated Reference: Digital Multimeter
Automated Reference: Fluke 5800A
LeCroy WR6100: (1 yr) CAL VER IEEE /5800+GHz+5ch,DMM
LeCroy WR6200: (1 yr) CAL VER IEEE /5800+GHz+5ch,DMM
Date: 2008-10-07 15:10:05
Revision: 1.1
Automated Reference: Digital Multimeter
Automated Reference: Fluke 5800A
Model(s) Covered:
WaveRunner 6100 and 6200
Source:
LeCroy WaveRunner® 6000 Series Service Manual, August 2004
The following tests have a Test Accuracy Ratio (TAR) of less than 4:1.
LEAKAGE VOLTAGE TEST (all channels)
DC, 50 Ohms : 20 mV/div, TAR 2:1
DC, 50 Ohms : 200 mV/div, TAR 2:1
DC, 1 MOhm : 20 mV/div, TAR 2:1
DC, 1 MOhm : 200 mV/div, TAR 2:1
DC, 1 MOhm : 2 V/div, TAR 2:1
DC, 50 Ohms : External, TAR 2:1
DC, 1 MOhm : External, TAR 2:1
Compatibility:
MET/CAL 7.20 or later
The average run time during testing was 16 minutes (2 channel),
31 minutes (4 channel).
Model(s) Covered:
WR6100 and 6200
Source:
LeCroy WaveRunner® 6000 Series Service Manual, August 2004
Compatibility:
MET/CAL 7.20 or later
The average run time during testing was 16 minutes (2 channel),
31 minutes (4 channel).