LeCroy WR6100: (1 yr) CAL VER IEEE /5800+GHz,DMM

LeCroy WR6200: (1 yr) CAL VER IEEE /5800+GHz,DMM

Date: 2008-10-07 15:09:59

Revision: 1.1

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5800A

Model(s) Covered:

WR6100 and 6200

Source:

LeCroy WaveRunner® 6000 Series Service Manual, August 2004

Compatibility:

MET/CAL 7.20 or later

The average run time during testing was 16 minutes (2 channel),

31 minutes (4 channel).

LeCroy WR6100: (1 yr) CAL VER IEEE /5800+GHz+5ch,DMM

LeCroy WR6200: (1 yr) CAL VER IEEE /5800+GHz+5ch,DMM

Date: 2008-10-07 15:10:05

Revision: 1.1

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5800A

Model(s) Covered:

WaveRunner 6100 and 6200

Source:

LeCroy WaveRunner® 6000 Series Service Manual, August 2004

The following tests have a Test Accuracy Ratio (TAR) of less than 4:1.

LEAKAGE VOLTAGE TEST (all channels)

DC, 50 Ohms : 20 mV/div, TAR 2:1

DC, 50 Ohms : 200 mV/div, TAR 2:1

DC, 1 MOhm : 20 mV/div, TAR 2:1

DC, 1 MOhm : 200 mV/div, TAR 2:1

DC, 1 MOhm : 2 V/div, TAR 2:1

DC, 50 Ohms : External, TAR 2:1

DC, 1 MOhm : External, TAR 2:1

Compatibility:

MET/CAL 7.20 or later

The average run time during testing was 16 minutes (2 channel),

31 minutes (4 channel).