Package: lem_unilap_geoX_1yr_ver_5320_vlc.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
LEM Unilap GeoX: (1 year) CAL VER /5320+VLC, Revision: 1.6
Norma Unilap GeoX: (1 year) CAL VER /5320+VLC, Revision: 1.6
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
LEM Unilap GeoX: (1 year) CAL VER /5320+VLC
Norma Unilap GeoX: (1 year) CAL VER /5320+VLC
Date: 2010-09-07 10:33:48
Revision: 1.6
Automated Reference: Fluke 5320A
Model Covered:
Unilap GeoX
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
DC LOW RESISTANCE (2 POLE)
1 Ohm, R1/R2; TUR 2.2000
10 Ohm, R1/R2; TUR 3.6667
DC LOW RESISTANCE (4 POLE)
1 Ohm, R1/R2; TUR 2.2000
10 Ohm, R1/R2; TUR 3.6667
AC RESISTANCE (2 POLE)
1 Ohm; TUR 2.2000
10 Ohm; TUR 3.6667
EARTH/GROUND RESISTANCE (3 POLE, Um = 48 V)
1 Ohm; TUR 2.2000
10 Ohm; TUR 3.6667
EARTH/GROUND RESISTANCE (4 POLE, Um = 20 V)
1 Ohm; TUR 2.2000
10 Ohm; TUR 3.6667
EARTH/GROUND RESISTANCE (4 POLE, R*, Um = 48 V)
1 Ohm; TUR 2.2000
10 Ohm; TUR 3.6667
Source:
Fluke Kassel Service Center Procedure
Fluke 1625 Earth/Ground Tester
Users Manual, January 2006
Source Author:
Fluke
Compatibility:
5500/CAL or MET/CAL 7.20 SP2 (7.20M) or later
Additional Equipment Required:
EI-1625, Selective/Stakeless Clamp Set for 1625
(Includes EI-162X, Clip-on Current Transformer
(sensing) with shielded cable set, EI-162AC,
Clip-on Current Transformer (inducing) and
2-3 wire adapter cable)
Lead Set, stackable, included in Fluke 5500/LEADS