Megger BM82/2: (1 yr) CAL VER /5320+VLC
Date: 2010-09-15 12:57:38
Revision: 1.1
Automated Reference: Fluke 5320A
Model(s) Covered:
BM82/2
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5320 (1 year) specifications.
INSULATION ACCURACY TESTS:
500 V, 3 GOhms; TUR 3.8667
1000 V, 9 GOhms; TUR 3.7037
1000 V, 50 MOhms; TUR 3.0000
1000 V, 90 MOhms; TUR 2.7778
1000 V, 500 MOhms; TUR 3.1250
1000 V, 900 MOhms; TUR 3.0028
1000 V, 3 GOhms; TUR 2.5128
Source:
Megger BM80/2
Instruction Manual
Specifications
Part no. 6172-188
Compatibility:
5500/CAL or MET/CAL 7.20Z or later
Additional Equipment Required:
None
The average run time during testing was minutes (MU mode).